Author(s): Berg, L.J.J. van den (2024)
Abstract:
Introduction
In the past decades, there has been a growing interest in incorporating glucagon into continuous
subcutaneous insulin infusion therapy for Type 1 Diabetes Mellitus patients. However,
widespread adoption as a standard treatment option has not yet been realized. The development
of bihormonal closed-loop systems involves investigating the pharmacodynamics of
glucagon in controlled settings. While some studies have examined these systems in real-life
settings, the focus has primarily been on their overall success in regulating glucose levels within
target ranges. Limited information is available on the real-life impact of glucagon on glucose
levels.
Objectives
This study aims to assess the effect of glucagon on glucose levels in real-life data, exploring
various conditions and investigating the impact of glucagon settings in the Inreda AP, the first
CE-certified bihormonal closed-loop system. The objective is to gain deeper insights into the
role of glucagon in glucose management, comprehend glucagon aspects within treatments with
the Inreda AP, and explore potential optimizations for enhanced effectiveness.
Methods
Glucose responses to glucagon are systematically mapped under various conditions, including
control (blockage), insulin on board, and different situations (baseline glucose value, rate of glucose
change and amount of glucagon) and patient characteristics (gender, age, BMI, diabetes
duration, HbA1c, total daily insulin dosages and insulin sensitivity). The impact of alterations in
system settings on regulatory outcomes is also examined. Cluster analysis is applied to predict
glucagon settings.
Conclusions
This research demonstrates that, in real-life scenarios, glucagon administration can lead to
earlier and more pronounced increases in glucose levels. The presence of insulin on board
mitigates glucose excursions after glucagon dosage, yet the pre-dosage rate of glucose change
appears to be a more decisive factor. Moreover, patients with lower insulin requirements exhibit
greater glucose excursions compared to those with higher insulin needs. Although there is
no one-size-fits-all glucagon setting, reducing settings results in decreased individual glucagon
usage. Practical recommendations for treatment adjustments and algorithm refinements are derived from this study, offering valuable insights to enhance the effective use of glucagon in glucose regulation.